The shortcomings of computer-aided measurement of migration for the prediction of failure of three forms of acetabular fixation by survival analysis and migration study to ten years

dc.authorid0000-0002-7426-0740
dc.contributor.authorScott, Gareth
dc.contributor.authorNakagawa, Shigeru
dc.contributor.authorOrhan, Zafer
dc.contributor.authorFreeman, Michael
dc.date.accessioned2021-06-23T19:18:51Z
dc.date.available2021-06-23T19:18:51Z
dc.date.issued2006
dc.departmentBAİBÜ, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümüen_US
dc.description.abstractWe studied, to ten years, migration by vertical displacement and angular change using a digitizer and standard radiographs and survival of three methods of acetabular fixation in primary hip replacement. Two implants were uncemented metal-backed components, one version of which was hydroxyapatite-coated. Both types had 28 mm polyethylene inserts. The third type was a cemented all-polyethylene cup with a 28 mm internal diameter All cups articulated with a Freeman neck-retaining stem. The purpose was to see if the early migration data predicted the long-term outcome and could be used to forecast implants that would fail for the purpose of pre-market assessment. No statistical differences were found between the groups in the parameters studied. At the threshold accuracies of our migration measurements (three millimetres and three degrees) at three years, we could not identify components that would subsequently fail. However, migration less than these threshold values and the absence of radiolucent lines were strongly associated with implant survival at ten years. Our system was insufficiently accurate for pre-market surveillance.en_US
dc.identifier.doi10.5301/HIP.2008.1277
dc.identifier.endpage249en_US
dc.identifier.issn1120-7000
dc.identifier.issue4en_US
dc.identifier.pmid19219800en_US
dc.identifier.scopus2-s2.0-33847692020en_US
dc.identifier.scopusqualityQ1en_US
dc.identifier.startpage243en_US
dc.identifier.urihttps://doi.org/10.1177/112070000601600401
dc.identifier.urihttps://hdl.handle.net/20.500.12491/5824
dc.identifier.urihttps://journals.sagepub.com/doi/pdf/10.1177/112070000601600401
dc.identifier.volume16en_US
dc.identifier.wosWOS:000243882500001en_US
dc.identifier.wosqualityN/Aen_US
dc.indekslendigikaynakWeb of Scienceen_US
dc.indekslendigikaynakScopusen_US
dc.indekslendigikaynakPubMeden_US
dc.institutionauthorOrhan, Zafer
dc.language.isoenen_US
dc.publisherWichtig Editoreen_US
dc.relation.ispartofHip Internationalen_US
dc.relation.publicationcategoryMakale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/closedAccessen_US
dc.subjectHip Arthroplastyen_US
dc.subjectAcetabular Componenten_US
dc.subjectMigrationen_US
dc.subjectSurvival Analysisen_US
dc.titleThe shortcomings of computer-aided measurement of migration for the prediction of failure of three forms of acetabular fixation by survival analysis and migration study to ten yearsen_US