The radiation hardness assurance facility at INFN-LNS catania for the irradiation of electronic components in air
Yükleniyor...
Dosyalar
Tarih
2010
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Ieee-Inst Electrical Electronics Engineers Inc
Erişim Hakkı
info:eu-repo/semantics/closedAccess
Özet
This paper describes the beam flux monitoring system that has been developed at the Superconducting Cyclotron at INFN-LNS (Istituto Nazionale di Fisica Nucleare, Laboratori Nazionali del Sud, Catania, Italy) in order to monitor the beam parameters such as energy, flux, beam profile, for SEE (Single Event Effects) cross-sections determination and DD (Displacement Damage) studies. In order to have an accurate and continuous monitoring of beam parameters we have developed fully automatic dosimetry setup to be used during SEE (with heavy ions) and DD (with protons up to 60 MeV) tests of electronic devices and systems. The final goal of our activity is to demonstrate that the operation of such a system in air is not detrimental to the accuracy on controlling the beam profile, energy and fluence delivered onto the DUT (Device Under Test) surface, even with non relativistic heavy ions. We have tested our beam monitoring system with the "Reference SEU monitor" developed by ESA/ESTEC, the results are discussed here.
Açıklama
10th European Conference on Radiation and Its Effects on Components and Systems (RADECS - 09) -- SEP 14-18, 2009 -- Bruges, BELGIUM
Anahtar Kelimeler
Dosimetry, Heavy Ion Accelerators, Radiation Effects, Radiation Tests
Kaynak
Ieee Transactions On Nuclear Science
WoS Q Değeri
Q1
Scopus Q Değeri
Q2
Cilt
57
Sayı
4