Evaluation of reliability of zirconia materials to be used in implant-retained restoration on the atrophic bone of the posterior maxilla: A finite element study

Yükleniyor...
Küçük Resim

Tarih

2019

Dergi Başlığı

Dergi ISSN

Cilt Başlığı

Yayıncı

Korean Acad Prosthodontics

Erişim Hakkı

info:eu-repo/semantics/openAccess

Özet

PURPOSE. Zirconia materials have been used for implant-retained restorations, but the stress distribution of zirconia is not entirely clear. The aim of this study is to evaluate the stress distribution and risky areas caused by the different design of zirconia restorations on the atrophic bone of the posterior maxilla. MATERIALS AND METHODS. An edentulous D4-type bone model was prepared from radiography of an atrophic posterior maxilla. Monolithic zirconia and zirconia-fused porcelain implant-retained restorations were designed as splinted or non-splinted. 300-N occlusal forces were applied obliquely. Stress analyses were performed using a 3D FEA program. RESULTS. According to stress analysis, the bone between the 1) molar implant and the 2) premolar in the non-splinted monolithic zirconia restoration model was stated as the riskiest area. Similarly, the maximum von Mises stress value was detected on the bone of the non-splinted monolithic zirconia models. CONCLUSION. Splinting of implant-retained restorations can be more critical for monolithic zirconia than zirconia fused to porcelain for the longevity of the bone.

Açıklama

9th International Biomechanic Congress -- SEP 19-22, 2018 -- Eskisehir, TURKEY

Anahtar Kelimeler

Finite Element Analysis, Dental Implant, Implant-retained Restoration, Zirconia, Posterior Maxilla

Kaynak

Journal Of Advanced Prosthodontics

WoS Q Değeri

Q3

Scopus Q Değeri

Q1

Cilt

11

Sayı

2

Künye