Temperature rises during tooth bleaching with different light-curing units
dc.authorid | 0000-0002-8637-5393 | en_US |
dc.authorid | 0000-0001-7346-3717 | en_US |
dc.authorid | 0000-0003-4802-1604 | |
dc.contributor.author | Özyılmaz, Özgün Yusuf | |
dc.contributor.author | Yavuz, Tevfik | |
dc.contributor.author | Sarı, Tuğrul | |
dc.contributor.author | Aykent, Filiz | |
dc.contributor.author | Öztürk, Atiye Nilgün | |
dc.date.accessioned | 2021-06-23T19:41:57Z | |
dc.date.available | 2021-06-23T19:41:57Z | |
dc.date.issued | 2015 | |
dc.department | BAİBÜ, Diş Hekimliği Fakültesi, Protetik Diş Tedavisi Ana Bilim Dalı | en_US |
dc.description.abstract | The aim of this study was to evaluate the influence of three different light-curing units on the temperature rise induced in the pulpal cavity, when bleached with 35% hydrogen peroxide (HP). Forty extracted maxillary central incisors were randomly divided into four groups according to photoactivation procedure (n=10). Whiteness HP (WHP) bleaching agent which contains 35% HP bleach was placed on the facial tooth surface and no photoactivation device was used (control); an blue light-emitting diode (LED, 40s), a diode laser (4W, Continous mode, 20s), and Nd:YAG laser (4W, 60Hz frequency, 20s). Temperature rise in the pulpal chamber was measured with a J-type thermocouple wire that was connected to a data logger before and immediately after each 20-40s application. One tooth of each group was selected for scanning electron microscope analysis. Data were analyzed using one-way ANOVA followed by the Games Howell post hoc test at a preset alpha of 0.05. One-way ANOVA revealed that there was a significant difference among the light-curing units (p<0.05) and between control group and light-curing units (p<0.05). The Nd:YAG induced significantly, which resulted in higher temperature increase than other light-curing units (10.7 degrees C). The LED unit produced the lowest temperature changes (3.2 degrees C) among the light-curing units. The presence of the bleaching gel did not cause a temperature increase in control group (0 degrees C). The results suggested that using light-curing units caused temperature changes in the intrapulpal cavity during tooth bleaching. The use of LED will be more safer for pulpal health when assisted with WHP bleaching gel. | en_US |
dc.identifier.doi | 10.1080/01694243.2015.1035490 | |
dc.identifier.endpage | 1580 | en_US |
dc.identifier.issn | 0169-4243 | |
dc.identifier.issn | 1568-5616 | |
dc.identifier.issue | 15 | en_US |
dc.identifier.scopus | 2-s2.0-84929516681 | en_US |
dc.identifier.scopusquality | Q2 | en_US |
dc.identifier.startpage | 1572 | en_US |
dc.identifier.uri | https://doi.org/10.1080/01694243.2015.1035490 | |
dc.identifier.uri | https://hdl.handle.net/20.500.12491/8256 | |
dc.identifier.volume | 29 | en_US |
dc.identifier.wos | WOS:000354532700005 | en_US |
dc.identifier.wosquality | Q3 | en_US |
dc.indekslendigikaynak | Web of Science | en_US |
dc.indekslendigikaynak | Scopus | en_US |
dc.institutionauthor | Yavuz, Tevfik | |
dc.language.iso | en | en_US |
dc.publisher | Taylor & Francis Ltd | en_US |
dc.relation.ispartof | Journal Of Adhesion Science And Technology | en_US |
dc.relation.publicationcategory | Makale - Uluslararası Hakemli Dergi - Kurum Öğretim Elemanı | en_US |
dc.rights | info:eu-repo/semantics/closedAccess | en_US |
dc.subject | Light-Curing Units | en_US |
dc.subject | Laser | en_US |
dc.subject | Thermocouple | en_US |
dc.subject | Bleaching | en_US |
dc.subject | Pulpal Health | en_US |
dc.title | Temperature rises during tooth bleaching with different light-curing units | en_US |
dc.type | Article | en_US |
Dosyalar
Orijinal paket
1 - 1 / 1
Küçük Resim Yok
- İsim:
- ozgun-yusuf-ozyilmaz.pdf
- Boyut:
- 483.65 KB
- Biçim:
- Adobe Portable Document Format
- Açıklama:
- Tam metin/Full text