Evaluation of Bio-Rad D-10® and arkray adams HA-8160® HPLC analyzers in HbA1c measurement
Dosyalar
Tarih
Dergi Başlığı
Dergi ISSN
Cilt Başlığı
Yayıncı
Erişim Hakkı
Özet
Glycosylated hemoglobin A1c (HbA1c) is an important parameter used for the assessment of time-dependent glycemic status and the diagnosis and follow-up of diabetes. This study aimed to evaluate the analytical performance of two HPLC analyzers, the Bio-Rad D-10®, and the Arkray Adams HA-8160®. Accuracy and imprecision studies were conducted, and a method comparison study was performed with 105 samples. Samples were collected on five consecutive days and measured on both analyzers within two hours. Bland-Altman and regression analyses were used for statistical evaluation of the data. Low and high-level within-day CV values were calculated as 1.22% and 0.60% for the Arkray analyzer and 1.2% and 0.30% for the Bio-Rad analyzer, respectively. They were calculated as 1.27% and 2.52% for the Arkray analyzer and 2.32% and 3.44% for the Bio-Rad analyzer, respectively, between days. The within-day CV values for both analyzers were below the limit of 2.5% specified by the International Federation of Clinical Chemistry (IFCC). The bias average for three months was 3.2% for the Arkray analyzer and 1.2% for the Bio-Rad analyzer. The Spearman correlation coefficient (r) was 0.973 (p<0.0001) between the two measurements, and the slope and intercept values calculated in the regression analyses showed the linear relationship and harmony between the two analyzers [R2 = 0.967, (y = 0.860x + 0.604)]. Both HPLC analyzers are shown to be reproducible and accurate for routine biochemistry laboratory use.