Üç farklı pakimetre ile ölçülen merkezi kornea kalınlığı ölçümlerinin değerlendirilmesi

dc.authorid0000-0003-4468-3985
dc.authorid0000-0002-8249-2621
dc.authorid0000-0002-3053-2396
dc.authorid0000-0001-8305-6600
dc.contributor.authorUlaş, Fatih
dc.contributor.authorDoğan, Ümit
dc.contributor.authorÇelik, Fatih
dc.contributor.authorSoydan, Adem
dc.contributor.authorÇelebi, Serdal
dc.date.accessioned2021-06-23T18:34:45Z
dc.date.available2021-06-23T18:34:45Z
dc.date.issued2014
dc.departmentBAİBÜ, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümüen_US
dc.description.abstractAmaç: Sağlıklı genç bireylerde ultrasonik pakimetre, speküler mikroskop ve temassız pnö- motonometre cihazına entegre optik pakimetre cihazlarıyla ölçülen merkezi kornea kalınlığı (MKK) sonuçlarının değerlendirilmesi. Gereç ve Yöntem: Bu kesitsel çalışmaya yaşları 21-32 (ortalama 24,54±2,71 yıl) arası de- ğişen 21 (%43,8) erkek ve 27 (%56,3) kadın sağlıklı bireyin sağ gözleri dahil edildi. Ultrasonik pakimetre, temassız speküler mikroskop cihazı ve pnömotonometreye entegre optik pakimetre cihazı ile ilk seans MKK ölçümleri birinci gözlemci tarafından ardışık iki ölçüm yapılarak alındı. İkinci seans MKK ölçümleri ise aynı cihazlarla birinci ve ikinci gözlemciler tarafından her gözlemci bir defa ölçüm yapacak şekilde yapıldı. Tüm MKK ölçümleri sabah saatlerinde, topikal anestezi altında ve en son ultrasonik pakimetre cihazı ile yapılmak üzere rastgele sı- rayla gerçekleştirildi. MKK ölçümleri ilişkili örneklem tek yönlü varyans analizi, sınıf içi ko- relasyon katsayısı, Bland-Altman grafiği, tekrarlanabilirlik katsayısı ve yeniden üretilebilirlik katsayısı ile değerlendirildi. Bulgular: Ultrasonik pakimetre, speküler mikroskop ve tonometreye entegre optik pakimetre cihazı ile ölçülen MKK değerlerinin ortalaması sırasıyla 549,84±31,46, 549,35±38,86 ve 539,71±27,81 ?m idi. Ultrasonik pakimetre, speküler mikroskop, tonometreye entegre optik pakimetre cihazları ile ölçülen değerlerin tekrarlanabilirlik ve yeniden üretilebilirlik katsayıları sırasıyla %2,72, %6,49, %4,55 ve %3,46, %5,80, %3,55 idi. Bland-Altman grafiğinde %95 güven aralığı ultrasonik pakimetre ve speküler mikroskop ölçümleri için yaklaşık 43 ?m, ultrasonik pakimetre ve tonometreye entegre optik pakimetre ölçümleri için yaklaşık 41 ?m gibi yüksek değerlerdeydi. Sonuç: Ultrasonik pakimetre ile yapılan MKK ölçümleri, speküler mikroskop ve tonometreye entegre optik pakimetre cihazlarıyla yapılan MKK ölçümlerinden daha yüksek tekrarlanabilirlik ve yeniden üretilebilirlik değerlerine sahiptir. Klinik uygulamalarda, bu optik pakimetrelerle elde edilen MKK ölçümleriyle ultrasonik pakimetrelerle ölçülen değerler birbirlerinin yerine kullanılmamalıdır.en_US
dc.description.abstractAim: To evaluate the central corneal thickness (CCT) measurements by ultrasonic pachymeter, non-contact specular microscope and non-contact pneumotonometer combined with an optical pachymeter. Material and Method: Right eyes of 21 (43.8%) male and 27 (56.3%) female healthy sub- jects aged between 21-32 years (mean 24.54±2.71) were included in this cross-sectional study. In the first session of CCT measurements the first operator took two consecutive measurements by ultrasonic pachymeter, non-contact specular microscope and non-contact pneumotonome- ter combined with an optical pachymeter. In the second session of CCT measurements the first operator took one more measurement and second operator took a further measurement using the same pachymeters. All CCT measurements were performed under topical anesthesia in the morning with random sequence but the ultrasound measurement was taken last. The CCT measurements were evaluated using one way analysis of variance, intraclass correlation coef- ficient, Bland-Altman plot, coefficient of repeatability and coefficient of reproducibility. Results: The mean CCT measurements using ultrasonic pachymeter, specular microscope and tonometer combined with pachymeter were 549.84±31.46, 549.35±38.86 and 539.71±27.81 μm, respectively. The coefficient of repeatability and coefficient of reproducibil- ity for measurements of ultrasonographic pachymeter, specular microscope, tonometer com- bined with pachymeter were 2.72%, 6.49%, 4.55% and 3.46%, 5.80%, 3.55%, respectively. In the Bland-Altman plot the 95% limits of agreement was approximately 43 μm for ultrasono- graphic pachymeter and specular microscope measurements and approximately 41 μm for ul- trasonographic pachymeter and tonometer combined with pachymeter measurements. Conclusion: Measurement of CCT with ultrasonographic pachymeter yielded better re- peatability and reproducibility than specular microscopy and tonometer combined with pachymeter. In clinical practice, the measurement acquired by these optical pachymeters are not interchangeable with ultrasonographic pachymeter measurements.en_US
dc.identifier.endpage87en_US
dc.identifier.issn1300-4786
dc.identifier.issue2en_US
dc.identifier.startpage81en_US
dc.identifier.trdizinid163833en_US
dc.identifier.urihttps://app.trdizin.gov.tr/makale/TVRZek9ETXpNdz09
dc.identifier.urihttps://hdl.handle.net/20.500.12491/2552
dc.identifier.volume21en_US
dc.indekslendigikaynakTR-Dizinen_US
dc.institutionauthorUlaş, Fatih
dc.institutionauthorDoğan, Ümit
dc.institutionauthorÇelik, Fatih
dc.institutionauthorSoydan, Adem
dc.institutionauthorÇelebi, Serdal
dc.language.isotren_US
dc.relation.ispartofMN Oftalmolojien_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectBland-Altman Grafiği
dc.subjectMerkezi Kornea Kalınlığı
dc.subjectPakimetre
dc.subjectTekrarlanabilirlik
dc.subjectYeniden Üretilebilirlik
dc.subjectBland-Altman Plot
dc.subjectCentral Corneal Thickness
dc.subjectPachymeter
dc.subjectRepeatibility
dc.subjectReproducibility
dc.titleÜç farklı pakimetre ile ölçülen merkezi kornea kalınlığı ölçümlerinin değerlendirilmesien_US
dc.title.alternativeEvaluation of central corneal thickness measurements with three different pachymetersen_US
dc.typeArticleen_US