Glokomlu olgularda dört farklı tonometreyle ölçülen göz içi basınç ölçüm sonuçlarının karşılaştırılması ve merkezi kornea kalınlığı ile ilişkileri

dc.authorid0000-0003-1018-925X
dc.authorid0000-0001-8305-6600
dc.authorid0000-0003-4468-3985
dc.authorid0000-0002-8249-2621
dc.contributor.authorKaymaz, Abdulgani
dc.contributor.authorÇelebi, Serdal
dc.contributor.authorUlaş, Fatih
dc.contributor.authorDoğan, Ümit
dc.date.accessioned2021-06-23T18:34:45Z
dc.date.available2021-06-23T18:34:45Z
dc.date.issued2014
dc.departmentBAİBÜ, Tıp Fakültesi, Cerrahi Tıp Bilimleri Bölümüen_US
dc.description.abstractAmaç: Glokomlu olgularda dört farklı tonometreyle ölçülen göz içi basınçlarının (GİB)karşılaştırılması ve merkezi kornea kalınlığı (MKK) ile ilişkilerinin araştırılması. Gereç ve Yöntem: Çalışmamızda glokom tanısı almış olguların Goldmann aplanasyon tonometresi (GAT), Tonopen tonometre (Tonopen Avia) ve nonkontakt tonometreler (NKT) (Canon TX-10 ve TX-20P) ile GİB ölçümleri ve ultrasonik pakimetre cihazı ile MKK ölçümleri yapıldı. Bulgular: Çalışmaya katılan 67 glokomlu olgunun yaş ortalaması 64,09±8,017 idi. Çalışmamızda Canon TX-10 ve TX-20P GİBstd (MKK'ye göre düzeltilmemiş Canon TX-20P GİB ölçümü) ile ölçülen GİB sonuçlarıyla MKK arasında istatistiksel olarak anlamlı orta düzeyde ilişki saptandı (sırasıyla TX-10 r=0,285, p=0,019, TX-20P GİBstd r=0,326, p=0,007). NKT TX-20P GİBcor (MKK'ye göre düzeltilmiş Canon TX-20P GİB ölçümü), Tonopen ve GAT ile ölçülen GİB değerleri ile MKK arasında istatistiksel olarak anlamlı ilişki saptanmadı (TX-20P GİBcor p=0,825, Tonopen p=0,980 ve GAT p=0,740). Canon TX-10 ve Tonopen ile ölçülen GİB değerleri GAT ile ölçülene göre hafif yüksek olmasına rağmen (sırasıyla 1,5 ve 1,8 mmHg) % 95’lik güven aralığı değerleri (sırasıyla 5,5 ve 5,6 mmHg) klinik olarak kabul edilemeyecek düzeylerdeydi. Canon TX-20P GİBstd ve TX-20P GİBcor değerleri GAT ile ölçülen GİB değerlerine göre klinik olarak kabul edilemeyecek düzeylerde yüksekti (sırasıyla 3 ve 3,5 mmHg). Sonuç: GAT, Canon TX-10, Canon TX-20P ve Tonopen ölçüm sonuçları birbirlerinin yerlerine kullanılmamalıdır.en_US
dc.description.abstractAim: To compare intraocular pressure measurements using four different tonometers and its relation with central corneal thickness in glaucoma patients. Material and Method: We measured IOP using Goldmann applanation tonometry (GAT), Tonopen tonometer (Tonopen Avia) and noncontact tonometer (NCT) (Canon TX-10 and TX-20P) and measured central corneal thickness (CCT) using ultrasonic pachymeter in glaucoma patients. Results: The study included 67 glaucoma patients with a mean age of 64.09±8.017. IOP values that were measured using NCT TX-10 and TX-20P IOPstd (IOP measurement of Canon TX-20P, which was not corrected according to central corneal thickness) had statistically moderate and significant correlation with CCT (respectively, TX-10 r=0.285, p=0.019, TX-20P IOPstd r=0.326 p=0.007). The IOP measurements of NCT TX-20P IOPcor (IOP measurement of Canon TX-20P, which was corrected according to central corneal thickness), Tonopen and GAT did not correlate significantly with CCT (TX-20P IOPcor p=0.825, Tonopen p=0.980 ve GAT p=0.740). Although IOP measurements of Canon TX-10 and Tonopen were slightly higher than GAT (1.5 mmHg and 1.8 mmHg, respectively), 95% confidence interval value had clinically unacceptable amplitude (5.5 mmHg and 5.6 mmHg, respectively). Canon TX-20PIOPstd and TX-20P IOPcor measurements were clinically unacceptably higher than GAT IOP measurements (3 mmHg and 3.5 mmHg, respectively). Conclusion: GAT, Canon TX-10, Canon TX-20P and Tonopen measurements should not be used interchangeably.en_US
dc.identifier.endpage252en_US
dc.identifier.issn1300-4786
dc.identifier.issue4en_US
dc.identifier.startpage245en_US
dc.identifier.trdizinid163774en_US
dc.identifier.urihttps://app.trdizin.gov.tr/makale/TVRZek56YzBOQT09
dc.identifier.urihttps://hdl.handle.net/20.500.12491/2551
dc.identifier.volume21en_US
dc.indekslendigikaynakTR-Dizinen_US
dc.institutionauthorKaymaz, Abdulgani
dc.institutionauthorÇelebi, Serdal
dc.institutionauthorUlaş, Fatih
dc.institutionauthorDoğan, Ümit
dc.language.isotren_US
dc.relation.ispartofMN Oftalmolojien_US
dc.relation.publicationcategoryMakale - Ulusal Hakemli Dergi - Kurum Öğretim Elemanıen_US
dc.rightsinfo:eu-repo/semantics/openAccessen_US
dc.subjectGlokom
dc.subjectGoldmann Aplanasyon Tonometresi
dc.subjectMerkezi Kornea Kalınlığı
dc.subjectNonkontakt Tonometre
dc.subjectTonopen Tonometre
dc.subjectGlaucoma
dc.subjectGoldmann Applanation Tonometer
dc.subjectCentral Corneal Thickness
dc.subjectNoncontact Tonometry
dc.subjectTonopen Tonometer
dc.titleGlokomlu olgularda dört farklı tonometreyle ölçülen göz içi basınç ölçüm sonuçlarının karşılaştırılması ve merkezi kornea kalınlığı ile ilişkilerien_US
dc.title.alternativeComparison of intraocular pressure measurements using four different tonometers and ıts relation with centrel corneal thickness in glaucoma patientsen_US
dc.typeArticleen_US