    Performance evaluation of urine osmolality measurement on Sysmex UF-5000 and the effect of molecules and particles in urine
    (Clinical Laboratory Publishing, 2023) Yis, Özgür Mehmet; Alışık, Murat; Buğdaycı, Güler; Sert, Mehmet A.; Erdoğan, Utku G.; Ateş, Merve
    Background: Urine osmolality determines the concentration ability of the kidney. Therefore, it is used to assess the body's hydration status, electrolyte levels, and acid-base disturbances. We aimed to evaluate the analytical perfor-mance of osmolality measurement of the Sysmex UF-5000 (UF-5000), to examine the effect of different molecules and particles in the urine on the osmolality measurement. Methods: Complete urinalysis and conductivity-based osmolality analysis with UF-5000 and osmolality analysis with Advanced (R) Model 3320 Micro-Osmometer (AI-3320) by freezing point reduction method were performed in the urine samples. Samples were grouped as negative, glucosuria, proteinuria, hematuria, pyuria, crystalluria, and urobilinogen.Results: Total impressions were calculated as < 5% and accuracy values were < 1.66% in both analyzers. The re-gression equation was found to be y =-12.54 + 0.956x and the relative difference between the analyzers was 8.7% in 586 samples. When patients with Glucose > 2+ were excluded the regression equation of 507 samples was found as y = 5.10 + 0.948x and the relative difference was 4.6%. The percentages of samples with a difference greater than the allowable difference were 18.8%, 11.6%, 35.9%, 13.7%, 18.7%, and < 12.2% in all samples, all samples without glucosuria > 2+, glucosuria, glucosuria < 3+, proteinuria, and other subgroups, respectively. Conclusions: Considering the good accessibility of the automated routine complete urine analyzer, UF-5000 can be considered to determine whether urine osmolality is within reference or should be measured by methods based on colligative properties. Thus, referral of patients to a clinic that uses the colligative measurement method may be used more effectively.(Clin. Lab. 2023;69:543-552. DOI: 10.7754/Clin.Lab.2022.220536)

| Bolu Abant İzzet Baysal Üniversitesi | Kütüphane | Rehber | OAI-PMH |

Bu site Creative Commons Alıntı-Gayri Ticari-Türetilemez 4.0 Uluslararası Lisansı ile korunmaktadır.


Bolu Abant İzzet Baysal Üniversitesi Kütüphanesi, Bolu, TÜRKİYE
İçerikte herhangi bir hata görürseniz lütfen bize bildirin

DSpace 7.6.1, Powered by İdeal DSpace

DSpace yazılımı telif hakkı © 2002-2025 LYRASIS

  • Çerez Ayarları
  • Gizlilik Politikası
  • Son Kullanıcı Sözleşmesi
  • Geri Bildirim